Advanced Operations

SYSTEM SETUP OUTPUT SETUP mA OUTPUT INT/EXT

Select OUTPUT card 1,2,3 or 4

 

 

 

Default value: No source

 

SELECT SOURCE

Selects a probe or creates a formula which delivers the process value that is processed by the

 

 

 

current output card.

 

 

 

 

 

SET PARAMETER

Default value: No Parameter

 

Selects a parameter of the selected source.

 

 

 

 

 

 

 

 

DATA VIEW

Default value: INPUT VALUE

 

Sets the displayed and logged measured value.

 

 

 

 

 

 

 

 

 

INPUT VALUE

Displays the process value which is read from the selected source after it has been processed

 

 

by the internal formula parser (if prompted)

 

 

 

 

 

CURRENT

Displays the calculated output current

 

 

 

 

 

SET FUNCTION

Default value: LINEAR CONTROL

 

 

 

 

 

 

LINEAR CONTROL

Tracks the measurement value.

 

 

 

 

 

 

PID CONTROL

Sets the sc1000 controller as a PID controller.

 

 

 

 

 

 

 

Default value: 10 mA

 

SET TRANSFER

Sets the substitute value for the output current in case the selected source reports an internal

 

 

 

error, is disconnected from the system or its output mode is set to 'Transfer value'.

 

ON ERROR MODE

Default value: SET TRANSFER

 

Sets the sc1000 controller reaction if an internal error occurs.

 

 

 

 

 

 

 

 

 

HOLD

The current output card operates constantly with the last valid value read from the selected

 

 

source.

 

 

 

 

 

SET TRANSFER

The current output card uses a substitute value for the output current.

 

 

 

 

 

SET MODE

Default value: DIRECT

 

Defines the point in time when the PID controller increases the Output current

 

 

 

 

 

 

 

 

 

DIRECT

The SNAP SHOT value is lower than the SETPOINT and vice versa.

 

 

 

 

 

 

REVERSE

The SNAP SHOT value is higher than the SETPOINT and vice versa.

 

 

 

 

 

SET FILTER

Sets the record time (in seconds)

 

The output current is based on the average of recorded values during a period of time. The

 

 

 

 

 

 

period set in this menu.

 

 

 

 

 

SCALE 0 mA/4 mA

Default value: 0-20 mA

 

Sets the output current range to 0–20 mA or 4–20 mA.

 

 

 

 

 

 

 

 

SET HIGH VALUE

Default value: 20

 

Sets the selected source value when the output current shall be 20 mA.

 

 

 

 

 

 

 

 

 

 

Default value: 0

 

SET LOW VALUE

Sets the selected source value when the output current shall be 0 mA (Scale is 0-20 mA)

 

 

 

respectively 4 mA (Scale is 4–20 mA).

 

 

 

Default value: 20 mA

 

MAXIMUM

Sets an upper limit for the possible output current value.

 

 

 

This menu item is displayed if SET FUNCTION is set to PID CONTROL.

 

 

 

 

 

 

 

Default value: 0 mA

 

MINIMUM

Sets the lower limit for the output current.

 

 

 

This menu item is displayed if SET FUNCTION is set to PID CONTROL.

 

 

 

 

 

 

 

Default value: 10

 

SET SETPOINT

Sets the process value

 

 

 

The PID controller tries to adjust to this process value.

 

 

 

 