Hach sc1000 Specifications, Plug-in Expansion Cards, DIN Rail Switch Cabinet Expansion Modules

Models: sc1000

1 150
Download 150 pages 30.78 Kb
Page 8
Image 8

Specifications

Plug-in Expansion Cards

Component description

Plug-in expansion cards for installation in the probe module

 

 

Operating temperature

–20 to 55 °C (–4 to 131 °F); 95% relative humidity, non-condensing

 

 

Storage temperature

–20 to 70 °C (–4 to 158 °F); 95% relative humidity, non-condensing

 

 

Analog output card

4 x analog current outputs (0–20 mA or 4–20 mA, max. 500 Ohm)

Terminals max. 1.5 mm2 (AWG15)

 

 

 

Analog/digital input card

4 x analog/digital inputs (0–20 mA or 4–20 mA) Terminals max. 1.5 mm2

 

4 x change over contacts (SPDT)

Internal relay card

Maximum switching voltage: 250 VAC, 125 VDC

Nominal Switching Current: 250 VAC, 5 A; 125 VAC, 5 A; 30 VDC, 5 A

 

 

Terminals max. 1.5 mm2 (AWG15)

 

 

Fieldbus interface card

Modbus RS485 (YAB021), Modbus RS232 (YAB047) or Profibus DP (YAB020)

 

 

DIN Rail Switch Cabinet Expansion Modules

 

 

Function

For installation in the switch cabinet. Any expansions required can be combined

when a base module is available.

 

 

 

Enclosure rating

IP20

 

 

Power supply

24 VDC (max. 30 V) from base module

 

 

Operating temperature

4 to 40 °C (39 to 104 °F); 95% relative humidity, non-condensing

 

 

Storage temperature

–20 to 70 °C (–4 to 158 °F); 95% relative humidity, non-condensing

 

 

 

Supply of expansion modules with 24 VDC and connection to the sc1000

 

network

 

Setting terminating resistor (with DIP switch) for the sc1000 network

Base module (LZX915)

Provision of connection for a display module (LXV402) for the configuration of

 

the system

 

Base module can supply a maximum of 2000 mA to the extension modules.

 

 

 

Dimensions: 23 x 100 x 115 mm (1 x 4 x 4.5 in.)

 

 

 

4 x normally closed into change over contacts (SPDT)

 

Maximum switching voltage: 250 VAC, 125 VDC

 

Maximum switching current: 250 VAC, 5 A; 125 VAC, 5 A; 30 VDC, 5 A

 

Maximum switching power: 150 W

Relay module (LZX920)

Can be programmed for limit, status-monitoring or for various control functions,

 

communication status indication by LED.

 

Terminals max. 2.5 mm2 (AWG 11)

 

Current consumption: <100 mA

 

 

 

Dimensions: 45 x 100 x 115 mm (2 x 4 x 4.5 in.)

 

 

 

2 analog current outputs (0–20 mA or 4–20 mA, max. 500 Ohm)

Output module (LZX919)

Terminals max. 2.5 mm2 (AWG 11)

Current consumption: <150 mA

 

 

 

 

Dimensions: 23 x 100 x 115 mm (1 x 4 x 4.5 in.)

 

 

 

Analog/Digital inputs (can be programmed as 0–20 mA or 4–20 mA), INPUT or

 

digital INPUT

Input module (LZX921)

Internal resistance: 180 Ohm

Terminals max. 2.5 mm2 (AWG 11)

 

 

Current consumption: <100 mA

 

 

 

Dimensions: 23 x 100 x 115 mm (1 x 4 x 4.5 in.)

 

 

6

Page 8
Image 8
Hach sc1000 user manual Specifications, Plug-in Expansion Cards, DIN Rail Switch Cabinet Expansion Modules