Standard Operations

 

Table 15

Error register

 

 

 

 

Bit

Error

 

Description

 

 

 

 

5

Hardware error

 

Hardware error detected.

 

 

 

 

6

Internal communication error

 

A communication failure within the device has been

 

detected.

 

 

 

 

 

 

 

7

Humidity error

 

Excessive humidity has been detected.

 

 

 

 

8

Temperature error

 

Temperature within the device exceeds specified limit.

 

 

 

 

10

Sample warning

 

Some action is required with the sample system.

 

 

 

 

11

Questionable calibration warning

 

The last calibration was of questionable accuracy.

 

 

 

 

12

Questionable measurement warning

 

One or more of the measurements of the device are of

 

questionable accuracy (Quality bad or out of range).

 

 

 

 

 

 

 

13

Safety warning

 

A condition has been detected which may result in a

 

safety hazard.

 

 

 

 

 

 

 

14

Reagent warning

 

Some action is required with the reagent system.

 

 

 

 

15

Maintenance required warning

 

Maintenance is required on this device.

 

 

 

 

Table 16 Status registerStatus 1

 

 

 

 

Bit

Status 1

 

Description

 

 

 

 

0

Calibration in progress

 

The device is in calibration mode. The measurements

 

may not be valid.

 

 

 

 

 

 

 

1

Cleaning in progress

 

The device is in cleaning mode. The measurements may

 

not be valid.

 

 

 

2

Service/Maintenance menu

 

The device is in service or maintenance mode. The

 

measurements may not be valid.

 

 

 

 

 

 

 

3

Common error

 

Device recognized an error, see Table 15 for more

 

detailed information

 

 

 

 

 

 

 

4

Measurement 0 Quality bad

 

Precision of measurement is out of specified limits.

 

 

 

 

5

Measurement low limit

 

Measurement is below specified range.

 

 

 

 

6

Measurement high limit

 

Measurement is above specified range.

 

 

 

 

7

Measurement 1Quality bad

 

Measurement is below specified range.

 

 

 

 

8

Measurement 1 low limit

 

Measurement is above specified range.

 

 

 

 

9

Measurement 1 high limit

 

Measurement is below specified range.

 

 

 

 

10

Measurement 2 Quality bad

 

Measurement is above specified range.

 

 

 

 

11

Measurement 2 low limit

 

Measurement is below specified range.

 

 

 

 

12

Measurement 2 high limit

 

Measurement is above specified range.

 

 

 

 

13

Measurement 3 Quality bad

 

Measurement is below specified range.

 

 

 

 

14

Measurement 3 low limit

 

Measurement is above specified range.

 

 

 

 

15

Measurement 3 high limit

 

Measurement is below specified range.

 

 

 

 

65