Advanced Operations
6.3.3.3 Function set to FEEDER CONTROL working mode
FEEDER CONTROL
|
|
| Default value: |
| SET TRANSFER | Sets the status of the relay | |
|
|
| the selected source or if the source is missing. |
|
|
|
|
| PHASE | Default value: HIGH | |
| Defines the relay status if the process value exceeds the set point. | ||
|
|
| |
|
|
|
|
|
| HIGH | Turns the relay on when the process value exceeds the set point. |
|
|
|
|
|
| LOW | Turns the relay on when the process value falls below the set point. |
|
|
|
|
| SET POINT | Default value: 10 | |
| Set the process value at which the relay toggles. | ||
|
|
| |
|
|
|
|
|
|
| Default value: 1 |
|
|
| Sets a hysteresis so the relay will not swing unregulated when the process value converges to |
| DEADBAND | the set point. | |
|
|
| PHASE is set to HIGH: The hysteresis is underneath the set point. |
|
|
| PHASE is set to LOW: The hysteresis is above the set point. |
|
|
|
|
|
|
| Default value: 0 minutes |
| OnMax TIMER | Sets a maximum period of time. During this period the relay is turned on when passing the set | |
| (0 | point. As soon as the time expires the relay is turned off regardless of the process value. | |
|
|
| 0=OnMax Timer is not active. |
|
|
| |
| ON DELAY | Default value: 5 seconds | |
| (0 | Sets a delay time for the relay to turn on. | |
|
|
| |
| OFF DELAY | Default value: 5 seconds | |
| (0 | Sets a delay time for the relay to turn off. | |
|
|
|
|