Advanced Operations

FREQ. Control /PID CONTROL

INPUT VALUE

Displays the process value read from the selected source after it has been processed by the internal formula parser (if prompted).

6.3.3.10 Function set to TIMER working mode

TIMER

 

 

 

Default value: No source

 

SENSOR

Select a probe or create a formula which delivers the process value that is to be processed by

 

 

 

the relay card.

 

 

 

 

 

 

 

Default value: NO

 

 

 

There is the opportunity to let the relay 'mark' the sensor configured in the menu SENSOR at the

 

 

 

DURATION time. Other sc1000 modules such as other relay cards or current output cards which

 

OUTPUTS ON HOLD

access data of this sensor read this 'mark' and go into hold. To go into hold means the accessing

 

module does not read the latest measurement from the marked sensor but works with the last

 

 

 

 

 

 

measurement read before the sensor was marked. To activate this function set this menu to

 

 

 

YES. If the sensor shall never let other modules go into hold, set this menu to NO.

 

 

 

Note: The setting OUTPUTS ON HOLD always adapts to the DURATION time.

 

 

 

 

 

 

YES

Adds a check mark to the selected SENSOR at the DURATION time. Other modules (relay card,

 

 

output card) accessing the probe go into hold mode as soon as they read the probe check mark.

 

 

 

 

 

NO

Sensor does not send other modules into hold mode

 

 

 

 

 

OFF DURATION

Default value: 30 seconds

 

Sets the period of time for turning off the relay in one duty cycle (provided the DUTY CYCLE

 

(0 s–65535 s)

 

option is set to DIRECT).

 

 

 

 

 

 

 

 

DURATION

Default value: 10 seconds

 

Sets the period of time for turning on the relay in one duty cycle (provided the DUTY CYCLE

 

(0 s–65535 s)

 

option is set to DIRECT).

 

 

 

 

 

 

 

 

 

 

Default value: 5 seconds

 

OFF DELAY

Delays the check mark of a probe even if DURATION time has expired.

 

The OFF DELAY time starts immediately after the DURATION time is expired.

 

(0 s–999 s)

 

This setting only takes effect if OUTPUTS ON HOLD is set to YES (see OUTPUTS ON HOLD

 

 

 

 

 

 

option).

 

 

 

 

DUTY CYCLE

Default value: DIRECT

 

 

 

 

 

 

DIRECT

Turns the relay ON for the time set in the DURATION menu.

 

 

Turns the relay OFF for the time set in the OFF DURATION menu.

 

 

 

 

 

 

 

 

 

REVERSE

Turns the relay OFF for the time set in the DURATION menu.

 

 

Turns the relay ON for the time set in the OFF DURATION menu.

 

 

 

 

 

 

 

 

INPUT VALUE

Displays the process value read from the selected source.

 

 

 

 

NEXT TOGGLE

Indicates the seconds until the relay will toggle.

 

 

 

 

 

 

 

Default value: OFF

 

LOG INTERVAL

Sets the interval for logging the displayed value into the data logger.

 

 

 

Options: OFF, 5min, 10min, 15min, 20min, 30min

 

 

 

 