Installation
3.6.3 Output card connections
If the instrument is equipped with the output card option, the mA output card supplies up to 4 analog
Note: The sc1000 mA output card cannot be used to provide power to a
To make an output card connection:
1.Remove power from the instrument. Remove the probe module cover.
2.Connect the output card to the appropriate slot (Figure 18). Use a magnetic screwdriver to secure the four screws to the card.
3.Install the card connector to the appropriate connection on the main circuit board (Figure 17).
4.Feed the cable through the base of the module and properly prepare and insert each wire into the terminal according to Figure 23 and Table 6. Tug gently after each insertion to make sure that the connection is secure.
5.Write the serial number from the rating plate on the supplied sticker and attach it to the main high voltage barrier (Figure 18).
6.Install the probe module cover.
After installation and connection of a
Figure 23 Output card (YAB019) cable connections
1Terminal
| Table 6 Output card (YAB019) terminal assignments | |
|
|
|
Terminal |
| Designation |
|
|
|
1 |
| Output 1+ |
|
|
|
2 |
| Output 1 – |
|
|
|
3 |
| Output 2 + |
|
|
|
4 |
| Output 2 – |
|
|
|
5 |
| Output 3 + |
|
|
|
6 |
| Output 3 – |
|
|
|
7 |
| Output 4 + |
|
|
|
8 |
| Output 4 – |
|
|
|
9 |
| Shield (Connected to protective earth) |
|
|
|
33