Advanced Operations

6.3.3.5 Function set to WARNING working mode

WARNING

 

 

Default value: Disabled

 

WARNING LIST

Sets the monitoring of the internal warning bits of the selected source.

 

ENABLED: Monitoring is active.

 

 

 

 

DISABLED: Monitoring is not active.

 

 

 

 

 

Default value: Disabled

 

ERROR LIST

Sets the monitoring of the internal error bits of the selected source.

 

ENABLED: Monitoring is active.

 

 

 

 

DISABLED: Monitoring is not active.

 

 

 

 

 

Default value: Disabled

 

PROCESS EVENT

Sets the monitoring of the internal process event bits of the selected source.

 

ENABLED: Monitoring is active.

 

 

 

 

DISABLED: Monitoring is not active.

 

 

 

 

 

Default value: DE-ENERGIZED

 

SET TRANSFER

Sets the status of the relay (ENERGIZED/DE-ENERGIZED) if some or all conditions (this means

 

warning, error or process event bits) are detected in the selected source or if the source is

 

 

 

 

missing.

 

 

 

 

ON DELAY

Default value: 5 seconds

 

(0 s999 s)

Sets a delay time for the relay to turn on.

 

 

 

 

OFF DELAY

Default value: 5 seconds

 

(0 s999 s)

Sets a delay time for the relay to turn off.

 

 

 