Remote Operation
CALCulate Subsystem Commands
CALCulate:MARKer[1234]:FUNCtion:BWIDthBANDwidth[:STATe] OFFON01
CALCulate:MARKer[1234]:FUNCtion:BWIDthBANDwidth[:STATe]?
Turns the bandwidth marker function ON or OFF for a particular marker. Only one marker function is allowed to be on at a time for each individual marker. Turning on the bandwidth function for a marker will turn off any other marker function and turn on the Bandwidth function. For example, turning on the Bandwidth function for a marker that has the Delta function ON, will turn OFF the Delta function and turn ON the Bandwidth function for the marker.
If the bandwidth function is turned ON for a marker that is OFF, the marker will be turned ON, placed at the center wavelength, and then the bandwidth markers will measure the bandwidth relative to this marker.
CALCulate:MARKer[1234]:FUNCtion:BWIDthBANDwidth:X:CENTer?
Returns the absolute
This query generates a “Settings conflict” error if the bandwidth function is OFF for the specified marker.
CALCulate:MARKer[1234]:FUNCtion:BWIDthBANDwidth:X:LEFT?
Returns the absolute
This query generates a “Settings conflict” error if the bandwidth function is OFF for the specified marker.