Using the Optical Spectrum Analyzer

The System Menus

Note

None of the Automeasure Setup parameters are affected by the front-panel Preset key.

They are not saved as part of the measurement setup.

To set the HP-IB address

1Press the front-panel System key.

2Press the More System Functions.... softkey.

3Press the Remote Setup.... softkey. The Remote Setup panel appears.

4The Current HPIB Address information shows the HP-IB address currently being used. To change the address, use the numerical entry pad or knob to change the New HPIB Address information. Press the Defaults softkey to reset the address to factory preset defaults. Refer to “To Fill In a Setup Panel” on page 3-65for information on changing and selecting items in the setup panel.

5Press the Set HPIB Address softkey if changes were made.

 

The Remote Setup panel

Setup panel

Current HPIB Address

selections

Shows the HP-IB address currently being used.

 

New HPIB Address

 

Enter the new HP-IB address in the text box using the number keys or knob.

3-55