Using the Optical Spectrum Analyzer
The System Menus
Setup panel selections
The Wavelength Calibration setup panel
Factory Wavelength Cal Date
Shows the date of the last
User Wavelength Cal Date
Shows the date of the last successful
Signal Source
Select either an external single wavelength signal source or the internal calibrator as the wavelength calibration source.
Wavelength Referenced In (for external calibration source only)
Selects air or vacuum for the calibration. If the internal calibrator is selected, this selection will not be available.
Set Calibration Wavelength (for external calibration source only)
This is the exact wavelength of the calibration source that will be used for the next user wavelength calibration. The wavelength entered must be within
2.5nm of the wavelength measured using the factory calibration. If the inter- nal calibrator is selected, this selection will not be available.
4If you are using an external calibration source, enter the signal source and wavelength of the calibration signal and whether the calibration is referenced in air or vacuum. If you are using an internal calibration source, continue with Step 5.