Table 5-9. OSA Errors (2 of 2)

Error Number

Error Description [description/explanation/examples]

 

 

20005

desc = “Trans-Impedance calibration data is invalid”

 

help = “Factory calibration data for the trans-impedance amplifier is invalid.

 

Please record the hexadecimal number listed with the error and cycle power. If

 

the error persists, contact the nearest Hewlett-Packard Instrument support

 

center for assistance. In the U.S., call (800) 403-0801. See the HP 86140 series

 

Users Guide for a listing of the HP sales and service offices.”

20006

desc = “Monochromator flatness calibration data is invalid”

 

help = “Factory flatness data for the monochromator is invalid. Please record the

 

hexadecimal number listed with the error and cycle power. If the error persists,

 

contact the nearest Hewlett-Packard Instrument support center for assistance.

 

In the U.S., call (800) 403-0801. See the HP 86140 series Users Guide for a

 

listing of the HP sales and service offices.“

20007

desc = “Sweep Timeout”

 

help = “A sweep was started but did not finish in the expected amount of time.

 

The trace data acquired during this sweep may not be valid. Try taking another

 

sweep. If the error persists, contact the nearest Hewlett-Packard Instrument

 

support center for assistance. In the U.S., call (800) 403-0801. See the HP 86140

 

series Users Guide for a listing of the HP sales and service offices.”

21999

desc = “The Error list has overflowed”

 

help = “The Error list has overflowed. The last entries received have been

 

deleted.”