Status Listings
OSA Warnings
Table
Error Number | Error Description [description/explanation/examples] |
|
|
5055 | desc = “Firmware Upgrade was not successful” |
| help => “A firmware upgrade operation was requested. The firmware upgrade |
| operation cannot be done at this time due to an internal software problem. “ + <- |
| .serviceCenterHelp |
5056 | desc = “Trajectory align cannot find input signal” |
| help = “The trajectory align procedure cannot find a usable input signal. Make |
| sure you have a signal connected to the optical input.” |
5057 | desc = “Invalid settings for trajectory align” |
| help = “The external trajectory align function was requested. It cannot be |
| executed because the start and/or stop wavelength settings are invalid. Start |
| and stop wavelengths must be between 600nm and 1700nm. In addition the |
| difference between start and stop wavelengths (span) must be greater than |
| 25nm. To correct this problem adjust the start and/or stop wavelength settings |
| to be valid.” |
5058 | desc = “Out of memory” |
| help = “The OSA has run out of execution memory. An internal function was |
| aborted due to lack of execution memory. The correction for this problem is to |
| cycle power.” |
5059 | desc = “Trajectory align: marginal input signal” |
| help = “The trajectory align procedure was aborted due to an input signal with |
| marginal amplitude. The input signal was large enough when the function began |
| it’s operation but later became too small. This usually happens when the signal |
| is on the edge of being too small. To correct this problem increase the input |
| signal level. If this is not possible then a slight increase in signal level can |
| sometimes be achieved by cleaning optical fiber connectors.” |
5060 | desc = “Trajectory align failed” |
| help = “The trajectory align procedure failed. The trajectory adjustments |
| computed were invalid. This is usually occurs after the OSA has received a large |
| mechanical shock. To correct this problem try the trajectory align procedure |
| again. If the error persists, contact the HP support center nearest your location.” |
5061 | desc = “Invalid marker trace” |
| help = “The marker could not be positioned. The marker is on a trace which does |
| not contain any valid data. This is sometimes caused by a trace math result |
| which is invalid.” |