Status Listings

OSA Warnings

Table 5-6. OSA Warnings (6 of 11)

Error Number

Error Description [description/explanation/examples]

 

 

5055

desc = “Firmware Upgrade was not successful”

 

help => “A firmware upgrade operation was requested. The firmware upgrade

 

operation cannot be done at this time due to an internal software problem. “ + <-

 

.serviceCenterHelp

5056

desc = “Trajectory align cannot find input signal”

 

help = “The trajectory align procedure cannot find a usable input signal. Make

 

sure you have a signal connected to the optical input.”

5057

desc = “Invalid settings for trajectory align”

 

help = “The external trajectory align function was requested. It cannot be

 

executed because the start and/or stop wavelength settings are invalid. Start

 

and stop wavelengths must be between 600nm and 1700nm. In addition the

 

difference between start and stop wavelengths (span) must be greater than

 

25nm. To correct this problem adjust the start and/or stop wavelength settings

 

to be valid.”

5058

desc = “Out of memory”

 

help = “The OSA has run out of execution memory. An internal function was

 

aborted due to lack of execution memory. The correction for this problem is to

 

cycle power.”

5059

desc = “Trajectory align: marginal input signal”

 

help = “The trajectory align procedure was aborted due to an input signal with

 

marginal amplitude. The input signal was large enough when the function began

 

it’s operation but later became too small. This usually happens when the signal

 

is on the edge of being too small. To correct this problem increase the input

 

signal level. If this is not possible then a slight increase in signal level can

 

sometimes be achieved by cleaning optical fiber connectors.”

5060

desc = “Trajectory align failed”

 

help = “The trajectory align procedure failed. The trajectory adjustments

 

computed were invalid. This is usually occurs after the OSA has received a large

 

mechanical shock. To correct this problem try the trajectory align procedure

 

again. If the error persists, contact the HP support center nearest your location.”

5061

desc = “Invalid marker trace”

 

help = “The marker could not be positioned. The marker is on a trace which does

 

not contain any valid data. This is sometimes caused by a trace math result

 

which is invalid.”

5-22