Using the Optical Spectrum Analyzer

The Markers Menus

To integrate between line markers

1Press the front-panel Markers key.

2Press the More Marker Functions.... softkey.

3Press the Line Marker Menu.... softkey.

4Select and set Wavelength Marker 1 and 2 to the desired integration limits.

5Press the Advanced Line Mkr Functions.... softkey.

6Press the Integrate Limit softkey to toggle this function ON or OFF.

Turning the integration limit ON will calculate the power between Wavelength Marker 1 and Wavelength Marker 2 when the Trace Integration function in ON. See “To calculate the total power under a trace” on page 3-28.

Note

Although there is a single range controlling the total power integration, the marker search range, and the wavelength sweep range, there are three independent state settings for limiting the total power integration, the marker search, and the wavelength sweep to the range.

To change the default marker settings

1Press the front-panel Markers key.

2Press the More Marker Functions.... softkey.

3Press the Marker Setup.... softkey.

4The Marker Setup panel opens. Refer to “To Fill In a Setup Panel” on page 3-65for information on changing and selecting items in the setup panel.

3-30