Remote Operation
DISPlay Subsystem Commands
DISPlay Subsystem Commands
DISPlay[:WINDow[1]]:ANNotation[:ALL] ONOFF01
DISPlay[:WINDow[1]]:ANNotation[:ALL]?
Turns the screen annotation ON or OFF. Affects only the
DISPlay[:WINDow[1]]:TEXT:CLEar
Erases all text on the display resulting from previous use of the DISPlay[:WIN- Dow[1]]:TEXT:DATA command.
DISPlay[:WINDow[1]]:TEXT:DATA <string><data_block>
DISPlay[:WINDow[1]]:TEXT:DATA?
Writes text on the display in the Title area. Use the <data_block> parameter to send extended ASCII characters such as control codes and symbols.
DISPlay[:WINDow[1]]:TRACe:ALL[:SCALe][:AUTO]
Finds the largest input signal using trace A and sets the span and vertical scale to display that signal. This command performs the same function as the front- panel
The following defines the instrument state settings altered by Auto Measure. State settings that are not listed are not altered.
Because many instrument state setting are altered, it is recommended you use this command only to find unknown signals. It is not recommend this command be used in the middle of a measurement routine.
Center Wavelength | According to signal wavelength and bandwidth |
Span | Set according to automeasure setup panel. In some cases, this |
| may also be a function of signal characteristics. |
Grating Order | Auto |
|
|