Using the Optical Spectrum Analyzer
The Amplitude Menus
Note
The optical spectrum analyzer performs a more complete zeroing when the instrument is first turned on. You can run this more complete routine anytime by pressing the Zero Now softkey. Refer to “To zero the instrument” on page
Auto Chop Mode
Enables the spectrum analyzer chop mode. Chop mode increases dynamic range for long sweep times by subtracting ambient light. Subtracting the ambient light can improve sensitivity but slow down the sweep speed.
Pressing the
Note
Sweep times of 50 msec × the # of trace points, or longer, use a special “chop” mode. Chop mode increases dynamic range, stabilizes measurements against drift, and reduces effects of stray light. This special chop mode is enabled by setting the Auto Chop mode to ON.
Power Calibration
Power Calibration allows you to use amplitude correction factors either from the factory calibration or from the last successful user calibration. Pressing the
User Power Cal Date
The User Power Cal Date displays the date of the last successful