Remote Operation

SENSe Subsystem Commands

[SENSe:][WAVelength:]SRANge[:STATe] OFFON01

[SENSe:][WAVelength:]SRANge[:STATe]?

Turns the wavelength sweep range ON or OFF. When the sweep range is ON, the instrument will only sweep between the upper and lower sweep range limits. There is a single range controlling the total power integration, the trace mean calculation, the marker search range, and the wavelength sweep range, but there are four independent state settings for limiting the total power integra- tion, the trace mean, the marker search, and the wavelength sweep to the range. If all four states for the total power integration, the trace mean, the marker search, and the wavelength sweep range are OFF, setting SENSe:WAVelength:SRANge:STATe to ON, will initialize

the lower limit to:

Start + S----pan---------and

 

 

 

3

 

the upper limit to:

Start +

2

S----pan---------

 

3

Sending the command when the instrument is in a zero span will generate a “Settings conflict” error.

[SENSe:][WAVelength:]SRANge:UPPer <numeric_value>[MNMUMAHZKHZMHZGHZ]

[SENSe:][WAVelength:]SRANge:UPPer?

Sets the upper limit for the wavelength sweep range. Setting this value when SENSe:WAVelength:SRANge:STATe is OFF will automatically turn SENSe:WAVelength:SRANge:STATe ON. The range used for the wavelength sweep range is the same range used for the total power integration, the trace mean range and the marker search range. Changing the range with this command will change all four ranges.

Sending the command when the instrument is in a zero span will generate a “Settings conflict” error. Default units for the parameter are meters. Frequency units are allowed.

4-87