Status Listings

Overview

Overview

 

This document describes the status listings of the HP 86140 series optical

 

spectrum analyzers. Status conditions for the optical spectrum analyzer are

 

grouped into categories.

Error

Indicates the instrument is malfunctioning. Measurement accuracy is probably

 

affected. Errors can be caused by either a hardware or a firmware problem.

 

The instrument requires repair at a Hewlett-Packard service center.

Status error

Indicates an internal hardware function is unavailable or not operating within

 

specifications. This is usually a temporary problem, such as a temperature

 

control loop being unsettled. Status errors cannot be cleared by the user.

 

When the condition causing the error is corrected the error will go away.

Warning

A warning is displayed when the optical spectrum analyzer cannot satisfy a

 

request from the user. Parameter entries that are out of range, illegal or unrec-

 

ognized remote commands or missing hardware options can cause a warning

 

to be displayed. A warning can be displayed if a requested measurement, for

 

example AutoMeasure, cannot be performed due to a missing input signal.

UNCAL

This message alerts the user the requested instrument setup can cause an

 

invalid measurement. For example, manually setting the sweep time to a value

 

that is too fast for other instrument settings.

Notice

A user alert indicating something in the state of the instrument has changed.

 

For example, if the user specifies a particular value for the current duty cycle,

 

a notice will be displayed when the optical spectrum analyzer modifies the

 

pulse width .

5-2