Using the Optical Spectrum Analyzer

The System Menus

The IP function is the same as the front-panel Preset key. For a complete list of the Preset conditions, see page 3-5. When Last is selected the instrument will power-on in the same state it was in when last powered-off.

To perform a factory preset (IP)

1Press the front-panel System key.

2Press the More System Functions.... softkey.

3Press the Service Menu.... softkey.

4Press the Factory Preset (IP) softkey.

The factory preset function restarts the instrument and sets it to the same states as the front-panel Preset key. See “SYSTem:PRESet” on page 4-93for a complete list of the preset conditions. In addition, it also sets the following functions.

Function

Factory Preset Value

 

 

Wavelength functions

 

 

 

User wavelength calibration date

None

 

 

User wavelength calibration source

None

 

 

User wavelength calibration wavelength

0 nm

 

 

Wavelength for next wavelength calibration

1150 nm

 

 

Wavelength for next amplitude calibration

1150 nm

 

 

User wavelength calibration correction

Zero

 

 

Amplitude functions

 

 

 

User amplitude calibration date

None

 

 

User amplitude calibration source

None

 

 

User amplitude calibration wavelength

0 nm

 

 

User amplitude calibration amplitude

–300 dBm

 

 

Amplitude for next amplitude calibration

0 dBm

 

 

User amplitude calibration correction

Zero

 

 

Auto Measure functions

 

 

 

Auto-Measure optimize sensitivity

Off

 

 

Auto-Measure auto span

On

 

 

Auto-Measure Manual span

10 nm

 

 

Auto-Measure dB/div auto

On

 

 

Auto-Measure manual dB/division

10 dB/division

 

 

3-49