Remote Operation
CALCulate Subsystem Commands
CALCulate[123456]:MEAN:RANGe:LOWer:TIME <numeric_value>[NSUSMSS]
Sets the lower
The range used for the trace mean range is the same range used for the total power calculation, the marker search range, and the wavelength sweep range. Changing the range with this command will change all four ranges. Default units for the parameters are seconds.
Sending this command while span is not set to zero will result in a “Settings Conflict” error.
CALCulate[123456]:MEAN:RANGe:LOWer[:WAVelength] <numeric_value>[MUMNMA]
This command sets the lower
The range used for the trace mean range is the same range used for the total power calculation, the marker search range, and the wavelength sweep range. Changing the range with this command will change all four ranges.
Sending this command when the instrument is in a zero span will generate a “Settings conflict” error. Default units for the parameter are meters. Frequency units are also allowed.
CALCulate[123456]:MEAN:RANGe[:STATe] OFFON01
CALCulate[123456]:MEAN:RANGe[:STATe]?
Turns the trace mean calculation range ON or OFF for all traces. Turning the calculation range ON will also turn the CALCulate:MEAN:STATe ON for the specified trace (the trace is specified via its subopcode). There is a single range controlling the total power integration, the trace mean range, the marker search range, and the wavelength sweep range, but there are four independent state settings for limiting the total power calculation, the trace mean calculation, the marker search, and the wavelength sweep to the range.