Reference
Cleaning Connections for Accurate Measurements
Table
Item | HP Part Number |
|
|
Isopropyl alcohol | |
Cotton swabs | |
Small foam swabs | |
Compressed dust remover | |
|
|
Table
Item | HP Part Number |
|
|
Laser shutter cap | |
FC/PC dust cap | |
Biconic dust cap | |
DIN dust cap | |
HMS10/HP dust cap | |
ST dust cap | |
|
|
| Cleaning a |
|
|
C A U T I O N | Do not use any type of foam swab to clean optical fiber ends. Foam swabs can |
| leave filmy deposits on fiber ends that can degrade performance. |
|
|
1Apply pure isopropyl alcohol to a clean
2Clean the ferrules and other parts of the connector while avoiding the end of the fiber.
3Apply isopropyl alcohol to a new clean
4Clean the fiber end with the swab or lens paper.
Do not scrub during this initial cleaning because grit can be caught in the swab and become a gouging element.