Remote Operation

SENSe Subsystem Commands

and

Span

Center = Stop -------------""""

2

With Wavelength Limit Off, the minimum value for the Center Wavelength is nominally

350.1nm. The maximum value is 1999.9 nm. These limits are valid for wavelengths ref- erenced in air or vacuum.

With Wavelength Limit On, the minimum value for the Center Wavelength is nominally

600.1nm. The maximum value is 1699.9 nm. These limits are valid for wavelengths ref- erenced in air or vacuum.

The Preset value for Wavelength Limit is On. The Preset value for Center Wavelength is 1150 nm.

[SENSe:][WAVelength:]CENTer:STEP:AUTO OFFON01

[SENSe:][WAVelength:]CENTer:STEP:AUTO?

When ON, the step size is automatic. When OFF, the step size is fixed. The value is set by the [SENSe:][WAVelength:]CENTer:STEP[:INCRement] com- mand.

[SENSe:][WAVelength:]CENTer:STEP[:INCRement] <numeric_value>[MNMUMA]

[SENSe:][WAVelength:]CENTer:STEP[:INCRement]?

Specifies the center wavelength step size.

[SENSe:][WAVelength:]LIMit OFFON01

[SENSe:][WAVelength:]LIMit?

Specifies whether the span is limited to the specified range of 600 to 1700 nm.

[SENSe:][WAVelength:]OFFSet <numeric_value>[MNMUMA]

[SENSe:][WAVelength:]OFFSet?

Specifies the wavelength offset. This is the offset between the measured wavelength and the displayed wavelength.

4-85