Using the Optical Spectrum Analyzer
The System Menus
To display the OSA Extended State information panel
1Press the
2Press the More System Functions.... softkey.
3Press the Service Menu.... softkey.
4Press the Adv Service Functions.... softkey.
5Press the More Adv Service Menu.... softkey.
6Press the OSA Extended State.... softkey. The OSA Extended State Information panel is displayed and can be printed.
The OSA Extended State function provides information useful to HP service personnel.
To change the automeasure defaults
1Press the
2Press the More System Functions.... softkey.
3Press the Auto Measure Setup.... softkey. The Automeasure Setup panel appears.