Status Listings
OSA Notices
OSA Notices
System control-related error messages or warnings
The OSA system changed a setting and generated a warning that the operation was performed.
Table
Error Number | Error Description [description/explanation/examples] |
|
|
1000 | desc = “Sensitivity forced to Auto” |
| help = “Sensitivity has been forced to the Auto setting because another |
| instrument setting was made which does not permit sensitivity to be in the |
| Manual setting.” |
1001 | desc = “Old Mezzanine board in system” |
| help = “This OSA has an old mezzanine board installed. It is not necessary to |
| upgrade the mezzanine board but the software will not be able to correctly |
| determine which light source options are available.” |
1002 | desc = “No Delta Marker Amplitude for“ |
| help = “The delta marker amplitude display has been turned off due to a units |
| mismatch. The amplitude units of the reference marker are different than those |
| of the delta marker. Values in differing units cannot be compared.” |
1003 | desc = “Trace length increased” |
| help = “Trace length has been increased because the wavelength limit function |
| was turned off. The minimum trace length is larger when the wavelength limit |
| function is off and trace length was below the new minimum value.” |
2999 | desc = “The Notices list has overflowed” |
| help = “The Notices list has overflowed. The last entries received have been |
| deleted.” |