Using the Optical Spectrum Analyzer
The Amplitude Menus
The Amplitude Menus
The Amplitude menus are accessed using the
Function | See.... |
|
|
Reference Level | |
|
|
Scale/Division | |
| |
|
|
Display Mode | |
| |
|
|
Sensitivity | |
|
|
Peak to Reference Level | |
|
|
Amplitude Setup.... | |
|
|
Note
Normally, the optical spectrum analyzer automatically selects the amplitude units according to whether the scale is logarithmic (dBm) or linear (watts). See “To change the default amplitude settings” on page
Linear scale
In linear scale, the graticule is scaled in linear units, proportional to the input power. The bottom of the graticule line represents zero watts. The top of the graticule represents the reference level. Therefore, with 10 graticule lines, each division represents 10% of the input signal at the top of the screen.
Logarithmic scale
Logarithmic scales range from 0.01 to 20.00