Remote Operation
SENSe Subsystem Commands
[SENSe:][WAVelength:]STOP <numeric_value>[MNMUMAHZKHZMHZGHZ]
[SENSe:][WAVelength:]STOP?
Specifies the stop wavelength. The center wavelength and span are adjusted so that:
Span
Start = Center –
2
and
Span
Stop = Center +
2
If the instrument is in zero span, this command sets the center wavelength to the value specified.
With Wavelength Limit Off, the minimum value for the Stop Wavelength is nominally
350.2nm. The maximum value is 2000 nm. These limits are valid for wavelengths refer- enced in air or vacuum.
With Wavelength Limit On, the minimum value for the Stop Wavelength is nominally
600.2nm. The maximum value is 1700 nm. These limits are valid for wavelengths refer- enced in air or vacuum.
The Preset value for Wavelength Limit is On. The Preset value for Stop Wavelength is 1700 nm.