Using the Optical Spectrum Analyzer

The Amplitude Menus

Note

The settings for sensitivity, video bandwidth and sweep time interact. If the sensitivity is set to manual, the video bandwidth and sweep time may be forced to Auto mode. If the video bandwidth is set to manual, the sensitivity and sweep time may be forced to Auto. If the sweep speed is set to manual and is set too fast, the OVER SWEEP indicator will come on in the display area. Since these settings interact, it is recommended that only one of the settings be changed, whichever setting is most important to the measurement task being performed.

To set the reference level to the peak value

1Press the front-panel Amplitude key.

2Press the Peak to REF LEVEL softkey.

The system sets the value of the reference level equal to the value of the high- est point on the active trace.

To change the default amplitude settings

1Press the front-panel Amplitude key.

2Press the Amplitude Setup.... softkey.

3The Amplitude Setup panel opens. Refer to “To Fill In a Setup Panel” on page 3-65for information on changing and selecting items in the setup panel.

The Amplitude Setup panel

3-11