Status Listings
Execution errors
An <error/event number> in the range
One of the following events has occurred:
•A <PROGRAM DATA> element following a header was evaluated by the device as outside of its legal input range or is otherwise inconsistent with the device’s capabilities.
•A valid program message could not be properly executed due to some device condition.
Execution errors shall be reported by the device after rounding and expres- sion evaluation operations have taken place. Rounding a numeric data ele- ment, for example, shall not be reported as an execution error. Events that generate execution errors shall not generate Command Errors,
Table
Error Number | Error Description [description/explanation/examples] |
|
|
desc = “Execution error” | |
| help = ““ |
| [This is the generic syntax error for devices that cannot detect more specific |
| errors. This code indicates only that an Execution Error as defined in IEEE 488.2, |
| 11.5.1.1.5 has occurred.] |
desc = “INIT:IMM command ignored” | |
| help = “An initiate immediate remote command was received. The instrument |
| was in continuous sweep mode when the command was received. The remote |
| command is ignored unless the instrument is in single sweep mode. To select |
| single sweep mode use the INIT:CONT OFF command.” |
desc = “Settings Conflict Error” | |
| help = “A request was made but the instrument settings resulting from the |
| request are in conflict with each other.” |
| [Indicates that a legal program data element was parsed but could not be |
| executed due to the current device state (see IEEE 488.2, 6.4.5.3 and 11.5.1.1.5.)] |