Remote Operation

CALCulate Subsystem Commands

CALCulate[123456]:MEAN:RANGe:UPPer?

This query returns the upper X-axis limit for the trace mean range calculation. The range used for the trace mean range is the same range used for the total power calculation, the marker search range, and the wavelength sweep range.

The returned value is in meters, unless the span is set to zero, in which case the returned value is in seconds.

CALCulate[123456]:MEAN:RANGe:UPPer:FREQuency <numeric_value>[HZKHZMHZGHZTHZ]

Sets the upper X-axis limit for the trace mean range calculation. Setting this value when CALCulate:MEAN:RANGe:STATe is OFF will automatically turn CALCulate:MEAN:RANGe:STATe ON. The range used for the trace mean range is the same range used for the total power calculation, the marker search range, and the wavelength sweep range. Changing the range with this command will change all four ranges. Default units for the parameter are Hertz.

CALCulate[123456]:MEAN:RANGe:UPPer:TIME <numeric_value>[NSUSMSS]

This command sets the upper limit for the trace mean range calculation. Setting this value when CALCulate:MEAN:RANGe:STATe is OFF will automatically turn CALCulate:MEAN:RANGe:STATe ON. The range used for the trace mean range is the same range used for the total power calculation, the marker search range, and the wavelength sweep range. Changing the range with this command will change all four ranges. Default units for the parameter are sec- onds.

Sending this command while span is not set to zero will result in a “Settings Conflict” error.

CALCulate[123456]:MEAN:RANGe:UPPer[:WAVelength] <numeric_value>[MUMNMA]

This command sets the upper X-axis limit for the trace mean range calcula- tion. Setting this value when CALCulate:MEAN:RANGe:STATe is OFF will automatically turn CALCulate:MEAN:RANGe:STATe ON. The range used for the trace mean range is the same range used for the total power calculation, the marker search range, and the wavelength sweep range. Changing the range with this command will change all four ranges. Default units for the parameter are meters.

Sending the command when the instrument is in a zero span will generate a “Settings conflict” error.