Remote Operation
TRIGger Subsystem Commands
TRIGger Subsystem Commands
TRIGger[:SEQuence]:DELay <numeric_value>[SMSUSNS]
TRIGger[:SEQuence]:DELay?
Specifies the trigger delay used to start a measurement.
TRIGger[:SEQuence]:OUTPut OFFONAUTO
TRIGger[:SEQuence]:OUTPut?
Controls the ADC trigger output. When OFF the signal will be a TTL LOW. When ON the signal will be a TTL HIGH. When AUTO is specified the signal will go HIGH before the sampling interval of the detector and go LOW after the sampling interval of the detector.
TRIGger[:SEQuence]:OUTPut:PULSe:DCYCle <numeric_value>
TRIGger[:SEQuence]:OUTPut:PULSe:DCYCle?
Sets the duty cycle of the sync output. This is equivalent to the
SOURce[n]:PULSe:DCYCle command.
TRIGger[:SEQuence]:OUTPut:PULSe:WIDTh <numeric_value>[SMSUSNS]
TRIGger[:SEQuence]:OUTPut:PULSe:WIDTh?
Sets the pulse width of the sync output. This is equivalent to the
SOURce[n]:PULSe:WIDTh command.
TRIGger[:SEQuence]:SLOPe POSitiveNEGativeEITHer
TRIGger[:SEQuence]:SLOPe?
Specifies the polarity of triggering used to start a measurement. Specifying the slope sets the trigger source to INTernal.