Using the Optical Spectrum Analyzer
The Markers Menus
To turn off all line makers and line marker functions
1Press the
2Press the More Marker Functions.... softkey.
3Press the Line Marker Menu.... softkey.
4Press the Line Markers Off softkey.
To sweep between line markers
1Press the
2Press the More Marker Functions.... softkey.
3Press the Line Marker Menu.... softkey.
4Select and set Wavelength Marker 1 and 2 to the desired wavelength sweep limits.
5Press the Advanced Line Mkr Functions.... softkey.
6Press the Sweep Limit softkey to toggle this function ON or OFF.
When the sweep limit is ON, the instrument will only sweep between Wave- length Marker 1 and Wavelength Marker 2.
To search between line markers
1Press the
2Press the More Marker Functions.... softkey.
3Press the Line Marker Menu.... softkey.
4Select and set Wavelength Marker 1 and 2 to the desired wavelength search limits.
5Press the Advanced Line Mkr Functions.... softkey.
6Press the Search Limit softkey to toggle this function ON or OFF.
When the search limit is ON, all the marker peak/pit searches will be between Wavelength Marker 1 and Wavelength Marker 2.