Status Listings

 

Application-Specific Warnings

Table 5-7. Application-Specific Warnings (4 of 6)

 

 

Error Number

Error Description [description/explanation/examples]

 

 

8021

desc = “The spec file could not be found.”

 

help = “The application tried to load a specification file which could not be found

 

in the internal memory. This may have occurred if the last file used by the

 

application was deleted. The filename is specified within the parentheses.”

8022

desc = “A closing quote is missing.”

 

help = “The specification file contained a line with an opening quote but no

 

closing quote could be found.”

8023

desc = “A default path name was used.”

 

help = “The imported specification file was given a default path name. All

 

specification files must have at least one PATH statement before any of the

 

measurement statements. A default path name was used either because some

 

measurement statements were used before a PATH statement or there was no

 

PATH statement at all in the file.”

8024

desc = “A variable is expected.”

 

help = “The statement expected a variable as an input parameter. The line

 

number indicates which statement did not have the proper input parameter

 

type.”

8025

desc = “The path name has already been used.”

 

help = “The application allows new names to be assigned to measurement

 

paths. Once assigned these names can no longer be used by other measurement

 

paths. Only the first thirty-two characters are significant. The line number

 

indicates where the problem was found. The word in the parentheses is the path

 

name causing the problem.”

8026

desc = “A path name was used where a variable is expected.”

 

help = “The input parameters for measurement statements can take variables as

 

inputs. These variables are the results of other measurement statements. A path

 

name is the name of a measurement path not the result of a measurement and

 

cannot be used as an input parameter. The input parameter was ignored for this

 

case. The line number indicates where the problem was found. The word in the

 

parentheses is the path name causing the problem.”

8027

desc = “A parameter value was outside the legal range.”

 

help = “A parameter value was used which was outside the allowable range.

 

The first number in the parentheses is the number of the parameter causing the

 

problem. The other two numbers are the minimum and maximum values

 

allowed. These values are in base units (e.g meters watts).”

5-31