Using the Optical Spectrum Analyzer

The System Menus

Active Function Area Assist

When this function is ON, the function set in the Active Function Area is automatically set to the first, or top, softkey function. For example, when you press the front-panel Bandwidth/Sweep key, the Active Function Area Assist function will set the Active Function Area to resolution bandwidth. This means you can immediately enter the desired resolution bandwidth. There is no need to press the Res BW softkey.

To set up a printer

1Press the front-panel System key.

2Press the Printer Setup.... softkey. The Printer Setup panel appears.

Setup panel selections

The Printer Setup panel

Printer Location

Selects either the internal printer or an external printer as the print destina- tion.

To perform a power calibration

1Press the front-panel System key.

2Press the Calibration.... softkey.

3Press the Power Cal Setup.... softkey. The Power Calibration Setup panel appears.

The Power Calibration setup panel

3-44