C A U T I O N
C A U T I O N
C A U T I O N
C A U T I O N
C A U T I O N
General Safety Considerations
Do not use too much liquid in cleaning the optical spectrum analyzer. Water can enter the
VENTILATION REQUIREMENTS: When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4° C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
Always use the
Do not connect ac power until you have verified the line voltage is correct as described in “Line Power Requirements” on page
This instrument has autoranging line voltage input. Be sure the supply voltage is within the specified range.
iv