Reference

Printer Head Cleaning Procedure

Procedure

C A U T I O N

Figure 6-1. Example of a static-safe workstation

1Turn off the HP 86140-series optical spectrum analyzer, and remove the line power cord.

2Place the instrument at a static-safe work station as described in the introduction to this procedure.

3Use a coin or screwdriver to open the printer door that is located on the top of the instrument.

Avoid dropping the coin or screwdriver, used to open the printer door, into the printer assembly.

4Lift up the paper latch as shown in the following diagram, and remove the paper.

6-8