Using the Optical Spectrum Analyzer
The System Menus
Setup panel selections
Factory Power Cal Date
Shows the date of the last
User Power Cal Date
Shows the date of the last successful
Set Calibration Power
This is the exact amplitude of the calibration source that will be used for the next power calibration. The power entered must be within +10 dB and
Set Calibration Wavelength
This is the approximate wavelength of the calibration source that will be used for the next user power calibration. The wavelength entered must be within 2.5 nm of the wavelength measured using the factory calibration.
4Enter the power and wavelength of the calibration signal.
5Press the Perform Calibration.... softkey.
6 Follow the
To perform a wavelength calibration
1Press the
2Press the Calibration.... softkey.
3Press the Wavelength Cal Setup.... softkey. The Wavelength Calibration Setup panel appears.