Specifications and Regulatory Information

Specifications

Specifications

 

HP 86140A and HP 86143A

HP 86142A and HP 86145A

 

 

 

WAVELENGTH

 

 

Range

600 nm to 1700 nm

600 nm to 1700 nm

Span Range (continuously variable)

0.2 nm to full range and zero span

0.2 nm to full range and zero span

Accuracy

 

 

After calibration with internal wavelength

± 0.025 nm (1510–1570),

± 0.025 nm (1510–1570),

reference signal a

± 0.035 nm (1570–1640)

± 0.035 nm (1570–1640)

After user calibration within ± 40 nm of

± 0.05 nm

± 0.05 nm

calibration signal a

 

 

After user calibration over full wavelength

± 0.2 nm

± 0.2 nm

range a

 

 

Absolute accuracy (2 year factory calibration

± 0.5 nm

± 0.5 nm

cycle) a

 

 

Reproducibility, 1 minute a

± 0.003 nm

± 0.003 nm

Span Linearity a,b

± 0.05 nm, for spans <40 nm

± 0.05 nm, for spans < 40 nm

Span Linearity (1525 to 1570 nm) a,b

± 0.02 nm

± 0.02 nm

Tuning Repeatability a

± 0.003 nm

± 0.003 nm

 

 

 

RESOLUTION BANDWIDTH (RBW)

 

 

FWHM (selectable) a,c

0.07, 0.1, 0.2, 0.5, 1, 2, 5, 10 nm

0.06, 0.1, 0.2, 0.5, 1, 2, 5, 10 nm

Corrected Resolution Bandwidth

 

 

Accuracy (using noise markers)a

 

 

0.5 nm, 1525 to 1610 nm

± 4%

± 3%

0.2 nm, 1525 to 1610 nm

± 6%

± 5%

0.1 nm, 1525 to 1610 nm

± 12%

± 10%

 

 

 

7-5