Operation

Power Supply Operation

Set V, I, and P The voltage, current and power setpoints can be limited to less than the supply rating Limits range to match the tolerance of connected equipment or any other criteria you may

have.

You can control the voltage, current and power setpoint limits through the SETPT LIMIT menu option. Once the limits have been changed from the supply’s default rated output, settings outside this range are no longer accepted.

Note Setpoint limits do not apply to the triggered outputs and auto sequence outputs.

To set the voltage limits:

Step #

Do This

You Will See

1

MENU

ERROR MSGS

 

 

 

 

 

2

until you see the “SETPT

SETPT LIMIT

 

LIMIT” option

 

3

ENTER

Voltage Limit

 

 

 

 

 

4

ENTER

High 0V

 

 

 

 

 

5

Use the Voltage knob, arrow keys,

High #####V

 

or numeric keypad to enter a value.

 

 

The value must be within 0V to

 

 

103% of the unit’s rated voltage.

 

6

ENTER

Low 0V

 

 

 

 

The Maximum setting is saved and the

 

 

Minimum setting appears.

7

Use the Voltage knob, arrow keys,

Low #####V

 

or numeric keypad to enter a value.

 

 

The value must be within 0V to

 

 

103% of the unit’s rated voltage.

 

8

ENTER

The setting is saved and the display

returns to its default operating mode.

 

 

 

 

 

100

Operating Manual for XDC Series Power Supply

Page 102
Image 102
Xantrex Technology XDC manual To set the voltage limits