Analog

Programming

Interface

0-10V Range

Calibration

Remote Interface Calibration Procedure

The 0-10V range of the analog programming interface must be calibrated separately. Follow the procedure exactly as for calibrating the 0-5V range, except that all the analog signals will be scaled by a factor of 2.

The commands will begin with the header:

CAL:ANAL:10V:...

The inputs to the programming lines should be approximately 1V for the minimum calibration level and 9V for the maximum calibration level.