Operation
Power Supply Operation
70 Operating Manual for XDC Series Power Supply
To set the Under-Power Protection:
Fold Protection options are:
None: Fold protection disabled
CC: Shutdown on entering CC mode
CV: Shutdown on entering CV mode
CP: Shutdown on entering CP mode
A programmable delay time causes the supply to wait before shutting down the
output.
Step # Do This You Will See
1OVP SET 0V
2 Press repeatedly until the
desired setting appears.
UPP SET 0W
3UPP SET 0W
4Use both the Current and Voltage
knobs, or the numeric keypad, or the
arrow keys to enter a value (must be
between 0W and 103% of the unit’s
rated power).
UPP SET ####W
5S/D if trip? N
6Use the arrow keys to select Yes or
No. For this example, Yes is selected. S/D if trip? Y
7This setting is saved and the display
returns to its default operating mode.
PROT SET
ALARMS
PROT SET
ALARMS
ENTER
ENTER
ENTER