Operation

Power Supply Operation

To set the Under-Power Protection:

Step #

Do This

You Will See

1

PROT SET

OVP SET 0V

ALARMS

2

PROT SET

 

ALARMS

Press repeatedly until the

UPP SET 0W

desired setting appears.

 

3

ENTER

UPP SET 0W

 

 

 

 

 

4

Use both the Current and Voltage

UPP SET ####W

knobs, or the numeric keypad, or the

 

arrow keys to enter a value (must be

 

 

between 0W and 103% of the unit’s

 

 

rated power).

 

5

ENTER

S/D if trip? N

 

 

 

 

 

6

Use the arrow keys to select Yes or

S/D if trip? Y

No. For this example, Yes is selected.

7

ENTER

This setting is saved and the display

 

 

returns to its default operating mode.

Fold Protection options are:

None: Fold protection disabled

CC: Shutdown on entering CC mode

CV: Shutdown on entering CV mode

CP: Shutdown on entering CP mode

A programmable delay time causes the supply to wait before shutting down the output.