SCPI Command Reference
SCPI Command Summary

Release 3.0 203

Table B.5Commands for CalibrationTable B.6Command to Clear all Protection Mechanisms
Function
SCPI Command
Description Query
Restore Factory
ion
[:]
CALibration[<channel>]:RESTore
Restonres the calibration to the constants set
at the factory
N/A
Change Calibration
Password
[:]
CALibration[<channel>][:SECure]:CODE <codeword>
Changes the calibration security code. No
Set Calibration State
[:]
CALibration[<channel>][:SECure]:STATe
<on-off-state>,<codeword>
Change calibration state (mode) Yes
Set Analog Current
Programming Input Level
[:]
CALibration[<channel>]:ANALog:<5V|10V>:PROGram:CURRe
nt:LEVel {MINimum|MAXimum}
Set analog programming current calibration
level
No
Enter Analog Current
Programming Input Data
[:]
CALibration[<channel>]:ANALog:
:{5V|10V}:PROGram:CURRent[:DATA] <current>
Set analog programming current calibration
data
No
Set Analog Voltage
Programming Input Level
[:]
CALibration[<channel>]:ANALog:
:{5V|10V}:PROGram:VOLTage:LEVel {MINimum|MAXimum}
Set analog programming voltage calibration
level
No
Enter Analog Voltage
Programming Input Data
[:]
CALibration[<channel>]:ANALog:
:{5V|10V}:PROGram:VOLTage[:DATA] <current>
Set analog programming voltage calibration
data
No
Set Analog Current
Readback Output Level
[:]
CALibration[<channel>]:ANALog:
:{5V|10V}:READback:CURRent:LEVel {MINimum|MAXimum}
Set analog readback current calibration level No
Enter Analog Current
Readback Output Data
[:]
CALibration[<channel>]:ANALog:
:{5V|10V}:READback:CURRent[:DATA] <current>
Set analog readback current calibration data No
Set Analog Voltage
Readback Output Level
[:]
CALibration[<channel>]:ANALog:
:{5V|10V}:READback:VOLTage:LEVel {MINimum|MAXimum}
Set analog readback voltage calibration level No
Enter Analog Voltage
Readback Output Data
[:]
CALibration[<channel>]:ANALog:
:{5V|10V}:READback:VOLTage[:DATA] <current>
Set analog readback voltage calibration data No
Set Supply Output Current
Level
[:]
CALibration[<channel>]:OUTPut:CURRent:LEVel
{MINimum|MAXimum}
Set output current calibration level No
Enter Output Current Data
[:]
CALibration[<channel>]:OUTPut:CURRent[:DATA]
<current>
Set output current calibration data No
Set Supply Output Voltage
Level
[:]
CALibration[<channel>]:OUTPut:VOLTage:LEVel
{MINimum|MAXimum}
Set voltage output calibration level No
Enter Output Voltage Data [:]CALibration[<channel>]:OUTPut:VOLTage[:DATA]
<voltage>
Set voltage output calibration data No
Function SCPI Command Description Query
Clear Output Protection [:]OUTPut[<channel>]:PROTection:CLEar Clears the protection mechanism. N/A