Calibration

Remote Interface Calibration Procedure

5.Enter voltage data

Enter the voltage at the voltage programming lines, read from the external DVM. CAL:ANAL:5V:PROG:VOLT:DATA <voltage>

6.Power supply calculates and stores calibration constants.

Analog Programming Interface Voltage Readback Calibration

1.Attach a DVM across the voltage readback lines, Pins B5 and B1(GND).

2.Minimum calibration level

Set the voltage readback lines to approximately 10% of full scale by sending the command:

CAL:ANAL:5V:READ:VOLT:LEV MIN

3.Enter voltage data

Enter the voltage across the voltage readback lines, read from the external DVM. CAL:ANAL:5V:READ:VOLT:DATA <voltage>

4.Maximum calibration level

Set the voltage readback lines to approximately 90% of full scale by sending the command:

CAL:ANAL:5V:READ:VOLT:LEV MAX

5.Enter voltage data

Enter the voltage across the voltage readback lines, read from the external DVM. CAL:ANAL:5V:READ:VOLT:DATA <voltage>

6.Power supply calculates and stores calibration constants.

Analog Programming Interface Current Programming Calibration

1.Connect the power source across the current programming lines, Pins B4 and B1(GND). Attach a DVM across the current programming lines as well.

2.Minimum calibration level

Set the input to the programming lines to approximately 0.5V (10% of full scale).

Set the power supply to receive 5V analog current programming calibration data with the command:

CAL:ANAL:5V:PROG:CURR:LEV MIN