Operation

Power Supply Operation

Aux line B has been selected to be configured.

To configure Aux line B:

Step #

Do This

You Will See

1

Press repeatedly until the

Cfg CV

 

desired option appears. For this

 

 

example, CV is selected.

 

2

ENTER

Pol Act High

 

 

 

3

Press repeatedly until the

Pol Act Low

 

desired option appears.

 

 

Select either “Act High” (Active high

 

 

logic) or “Act Low” (Active low logic)

 

 

for the auxiliary lines. In this example,

 

 

Active Low is selected.

 

 

 

 

4

ENTER

This setting is saved and the display

returns to its default operating mode.

 

 

Configure

Power ON

Settings

The Power ON configuration can be set with 4 options:

Factory Preset: Where the Power ON output is reset to the original factory levels.

These include: Output=OFF, VSET=0, and ISET=0. (default configuration)

Last Setting: Where the Power ON output is set to the same level as when it was last powered OFF. This is useful for automatic recovery from short power failures.

User Settings: Where the Power ON output is set to a stored setting that is recalled from memory. See “Store User Settings” on page 77.

Auto Sequence: Where the Power ON output can be recalled from memory. See “To edit the sequence’s trigger source:Using Auto Sequencing” on page 96.

The output state also depends on the Power ON Output setting. The values listed above are true only if the OUT ON? setting is Yes. Otherwise Output is off (unit in Standby mode).

See “Recall Settings” on page 79 for settings affected by the Power ON feature.

Each of the 4 Power ON configuration options can be accessed from the Front Panel menus or remotely through a set of digital commands. The procedures that follow show how to use the Front Panel menus.