Remote Operation
SCPI Commands for Digital Interfaces
130 Operating Manual for XDC Series Power Supply
Configure V, I,
P Protection
Limits
Over-Voltage Protection
To set the Over-Voltage Protection level:
SOUR:VOLT:PROT <voltage>
To check the Over-Voltage Protection level:
SOUR:VOLT:PROT?
To check if the Over-Voltage Protection was tripped:
SOUR:VOLT:PROT TRIP?
Alternatively, you can query the status registers. See Appendix B.
Under-Voltage Protection
To set the Under-Voltage Protection level:
SOUR:VOLT:PROT:UND <voltage>
SOUR:VOLT:PROT:UND:STAT <on-off-state>
The first UVP protection command sets the protection level. The second command
lets you choose a warning alarm only [OFF] or shut down with an alarm [ON] if the
protection level is exceeded.
To check the Under-Voltage Protection level:
SOUR:VOLT:PROT:UND?
To check if the Under-Voltage Protection was tripped:
SOUR:VOLT:PROT:UND:TRIP?
Alternatively, you can query the status register. See AppendixB, “SCPI Command
Reference”.
Over-Current Protection
T
o set the Over-Current Protection level:
SOUR:CURR:PROT <current>
SOUR:CURR:PROT:STAT <on-off-state>
The first OCP protection command sets the protection level. The second command
lets you choose a warning alarm only [OFF] or shut down with an alarm [ON] if the
protection level is exceeded.
To check the Over-Current Protection level:
SOUR:CURR:PROT?
To check if the Over-Current Protection was tripped:
SOUR:CURR:PROT:TRIP?