Calibration

Remote Interface Calibration Procedure

3.Enter voltage data

Enter the voltage at the current programming lines, read from the external DVM. CAL:ANAL:5V:PROG:CURR:DATA <voltage>

4.Maximum calibration level

Set the input to the programming lines to approximately 4.5V (90% of full scale).

Set the power supply to receive 5V analog current programming calibration data by sending the command:

CAL:ANAL:5V:PROG:CURR:LEV MAX

5.Enter voltage data

Enter the voltage at the current programming lines, read from the external DVM. CAL:ANAL:5V:PROG:CURR:DATA <voltage>

6.Power supply calculates and stores calibration constants.

Analog Programming Interface Current Readback Calibration

1.Attach a DVM across the current readback lines, Pins B6 and B1(GND).

2.Minimum calibration level

Set the current readback lines to approximately 10% of full scale by sending the command:

CAL:ANAL:5V:READ:CURR:LEV MIN

3.Enter voltage data

Enter the voltage across the current readback lines, read from the external DVM. CAL:ANAL:5V:READ:CURR:DATA <voltage>

4.Maximum calibration level

Set the current readback lines to approximately 90% of full scale by sending the command:

CAL:ANAL:5V:READ:CURR:LEV MAX

5.Enter voltage data

Enter the voltage across the current readback lines, read from the external DVM. CAL:ANAL:5V:READ:CURR:DATA <voltage>

6.Power supply calculates and stores calibration constants.