Performance Tests

b.Set oscilloscope controls:

Vertical

CH1

Coupling

DC

Scale

5008mV/div.

Input impedance

508Ω

Horizontal

 

Sweep

5008ns/div.

Trigger

 

Source

CH1

Coupling

DC

Slope

Positive

Level

5008mV

Mode

Auto

2.Set the AWG2005 controls and select the waveform file:

a.Initialize AWG2005 controls: Push UTILITY￿Misc￿Config...￿Re/ set to Factory￿O.K.

b.Select the waveform file:

￿Push SETUP￿Waveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).

￿Turn the general purpose knob to select the CLK_￿MPL.WFM file.

￿Push ENTER to select the file.

3.Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.

4.Check clock amplitude accuracy: Check that the pulse amplitude of the displayed waveform is 2 Vp9p or greater.

5.End procedure: Disconnect the oscilloscope.

Gain ￿ccuracy Check This procedure checks the accuracy of the AWG2005 gain.

Electrical Characteristic Checked: Main Output, Amplitude, DC Accu9

racy, on page81912.

Equipment Required: A 508Ω coaxial cable, a 50 Ω termination, a BNC9to9 dual banana adapter, and a digital multimeter (DMM).

Prerequisites: The AWG2005 meets the prerequisites listed on page 497.

4￿30

Performance Verification