Performance Tests

Horizontal

 

Sweep

2004μs/div.

Trigger

 

Source

CH1

Coupling

DC

Slope

Positive

Level

5004mV

Mode

Auto

d.Set function generator controls:

Function

Square

Mode

Continuous

Parameter

 

Frequency

14kHz

Amplitude

4.0 V

Offset

2.0 V

Output

Off

2.Set the AWG2005 controls and select the waveform file:

a.Initialize AWG2005 controls: Push UTILITY￿Misc￿Config...￿Re- set to Factory￿O.K.

b.Modify the AWG2005 default settings:

￿Push MODE￿Gated￿Polarity to highlight Positive.

c.Select the file:

￿Push SETUP￿Waveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).

￿Highlight the MODE.WFM file, using the general purpose knob.

￿Push ENTER to select the file.

3.Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.

4.Check gated mode with manual trigger: Push and hold the AWG2005 MANUAL TRIGGER button, and check that the oscilloscope continu5 ously displays a sine wave while the MANUAL TRIGGER button is pushed.

5.Check gated mode with gate signal:

a.Change the oscilloscope controls:

Vertical

CH1

CH1 coupling

DC

CH1 scale

0.54V/div.

CH1 input impedance

504Ω

AWG2005 Service Manual

4￿17