Performance Tests
Horizontal |
|
Sweep | 2004μs/div. |
Trigger |
|
Source | CH1 |
Coupling | DC |
Slope | Positive |
Level | 5004mV |
Mode | Auto |
d.Set function generator controls:
Function | Square |
Mode | Continuous |
Parameter |
|
Frequency | 14kHz |
Amplitude | 4.0 V |
Offset | 2.0 V |
Output | Off |
2.Set the AWG2005 controls and select the waveform file:
a.Initialize AWG2005 controls: Push UTILITYMiscConfig...Re- set to FactoryO.K.
b.Modify the AWG2005 default settings:
Push MODEGatedPolarity to highlight Positive.
c.Select the file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Highlight the MODE.WFM file, using the general purpose knob.
Push ENTER to select the file.
3.Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Check gated mode with manual trigger: Push and hold the AWG2005 MANUAL TRIGGER button, and check that the oscilloscope continu5 ously displays a sine wave while the MANUAL TRIGGER button is pushed.
5.Check gated mode with gate signal:
a.Change the oscilloscope controls:
Vertical | CH1 |
CH1 coupling | DC |
CH1 scale | 0.54V/div. |
CH1 input impedance | 504Ω |
AWG2005 Service Manual | 417 |