Performance Tests

b.Set function generator controls:

Function

Square

Mode

Continuous

Parameter

 

Frequency

16kHz

Amplitude

06V

Offset

5.06V6

(The actual voltage out is 1/2 of what is displayed.)

Output

Off

2.Enable the function generator output: Turn on the function generator output.

3.Check external ADD operation: a. Check Autostep Step 4:

￿Push the AWG2005 MANUAL TRI￿￿ER button, and check that the step changes to Step 4 on the MODE menu.

￿Check that the DMM reading is in the range from 4.750 to 5.2506V.

4.Disable function generator output: Turn the function generator output off.

5.End procedure: Keep the test connections and instrument settings for the next check.

Check Internal Add Operation

Electrical Characteristic Checked: Arithmetic Operation, Add, on

page61713.

Equipment Required: Two 506Ω coaxial cables, a function generator, and a digital multimeter (DMM).

Prerequisites: The AWG2005 meets the prerequisites listed on page 477.

Procedure:

1.Use test hookup and control settings from previous check.

2.Check internal ADD operation: a. Check Autostep Step 5:

￿Push the AWG2005 MANUAL TRI￿￿ER button, and check that the step changes to Step 5 on the MODE menu.

￿Check that the DMM reading is in the range from 4.750 to 5.2506V.

4￿26

Performance Verification