Performance Tests

2.Set the AWG2005 controls and select the waveform file:

a.Initialize AWG2005 controls: Push UTILITY￿Misc￿Config...￿Re. set to Factory￿O.K.

b.Select waveform file:

￿Push SETUP￿Waveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).

￿Turn the general purpose knob to select the PULSE.WFM file.

￿Push ENTER to select the file.

￿Select the PULSE.WFM file for CH2 same as CH1.

3.Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.

4.Check pulse response at 0.5*V amplitude:

a.Check rise time: Check that the rise time of the waveform displayed on the oscilloscope from the 10% point to the 90% point is 354ns or less.

b.Check aberrations: Check that the aberrations of the displayed waveform is within 0.354div.

c.Check flatness: Check that the flatness of the displayed waveform is within 0.154div. after 204ns from the rising edge.

d.Change the oscilloscope controls:

Trigger

SlopeNegative

e.Check fall time: Check that the fall time of the displayed waveform from the 10% point to the 90% point is 354ns or less.

5.Check pulse response at 10*V amplitude:

a.Change the oscilloscope controls:

Vertical

CH1

CH1 scale

24V/div.

Trigger

 

Slope

Positive

b.Change the AWG2005 controls:

￿Push SETUP￿Amplitude to change the amplitude for CH1.

￿Press the numeric key 1, 0 and press the units key V to select an amplitude of 104V.

c.Repeat substeps 4a through 4e, checking to the follow limits:

Rise time

354ns, maximum

Aberrations

0.354div., maximum

AWG2005 Service Manual

4￿35