Performance Tests
2.Set the AWG2005 controls and select the waveform file:
a.Initialize AWG2005 controls: Push UTILITYMiscConfig...Re. set to FactoryO.K.
b.Select waveform file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to select the PULSE.WFM file.
Push ENTER to select the file.
Select the PULSE.WFM file for CH2 same as CH1.
3.Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Check pulse response at 0.5*V amplitude:
a.Check rise time: Check that the rise time of the waveform displayed on the oscilloscope from the 10% point to the 90% point is 354ns or less.
b.Check aberrations: Check that the aberrations of the displayed waveform is within 0.354div.
c.Check flatness: Check that the flatness of the displayed waveform is within 0.154div. after 204ns from the rising edge.
d.Change the oscilloscope controls:
Trigger
SlopeNegative
e.Check fall time: Check that the fall time of the displayed waveform from the 10% point to the 90% point is 354ns or less.
5.Check pulse response at 10*V amplitude:
a.Change the oscilloscope controls:
Vertical | CH1 |
CH1 scale | 24V/div. |
Trigger |
|
Slope | Positive |
b.Change the AWG2005 controls:
Push SETUPAmplitude to change the amplitude for CH1.
Press the numeric key 1, 0 and press the units key V to select an amplitude of 104V.
c.Repeat substeps 4a through 4e, checking to the follow limits:
Rise time | 354ns, maximum |
Aberrations | 0.354div., maximum |
AWG2005 Service Manual | 435 |