Performance Tests
b.Select waveform file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to highlight the CNTRL.WFM file.
Push ENTER to select the file.
3.Check rear$panel CONTROL SIG OUT amplitude:
a.Check CONTROL SIG OUT pulse amplitude:
Push ModeGated.
Push the AWG2005 MANUAL TRIGGER button.
Check that the pulse upper level of the displayed waveform is 26V or greater.
Check that the pulse lower level of the displayed waveform is 0.86V or less.
4.nd procedure: Disconnect the oscilloscope.
External Trigger Level This procedure checks the external trigger level accuracy of the AWG2005.
Accuracy CheckElectrical Characteristic Checked: Auxiliary Input, TRIGGER, Accuracy, on page61715.
Equipment Required: Two 506Ω coaxial cables, a function generator, and an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 477.
Procedure:
1.Install test hookup and set test equipment controls:
a.Hook up oscilloscope: Connect the AWG2005 CH1 output through a coaxial cable to the oscilloscope CH1 vertical input.
b.Hook up function generator: Connect the AWG2005 TRIGGER INPUT through a coaxial cable to the function generator output (see Fig7 ure4718).
AWG2005 Service Manual | 439 |