Performance Tests
Trigger
SourceCH1
b.Apply gate signal: Turn function generator output on.
c.Check gated mode with positive gate signal: Check that the oscillo0 scope displays a sine wave while the function generator gate signal level is in upper portion of the display (see Figure 406).
Waveform
Output
CH1
Gate
Signal
CH2
Figure
Waveform Output Signal
d.Change the AWG2005 trigger polarity to negative: Push MODEPo- larity to change the polarity to Negative.
e.Check gated mode with a negative gate signal: Check that the oscilloscope displays a sine wave while the function generator gate signal level is in the lower portion of the display.
6.End procedure: Turn the function generator output off and disconnect the function generator.
Check Waveform Advance Mode
Electrical Characteristic Checked: Operating modes, Waveform Ad0
vance, on page/105.
Equipment Required: A 50/Ω coaxial cable and an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 407.
418 | Performance Verification |