Performance Tests

Trigger

SourceCH1

b.Apply gate signal: Turn function generator output on.

c.Check gated mode with positive gate signal: Check that the oscillo0 scope displays a sine wave while the function generator gate signal level is in upper portion of the display (see Figure 406).

Waveform

Output

CH1

Gate

Signal

CH2

Figure 4-6:,Relationship between 1 Volt or Greater Gate Signal and

Waveform Output Signal

d.Change the AWG2005 trigger polarity to negative: Push MODE￿Po- larity to change the polarity to Negative.

e.Check gated mode with a negative gate signal: Check that the oscilloscope displays a sine wave while the function generator gate signal level is in the lower portion of the display.

6.End procedure: Turn the function generator output off and disconnect the function generator.

Check Waveform Advance Mode

Electrical Characteristic Checked: Operating modes, Waveform Ad0

vance, on page/105.

Equipment Required: A 50/Ω coaxial cable and an oscilloscope.

Prerequisites: The AWG2005 meets the prerequisites listed on page 407.

4￿18

Performance Verification